Assessing zinc from a nail clipping using mono-energetic portable X-ray fluorescence

dc.contributor.authorFleming, David E.B.
dc.contributor.authorCrook, Samantha L.
dc.contributor.authorEvans, Colby T.
dc.date.accessioned2025-01-09T19:29:10Z
dc.date.available2025-01-09T19:29:10Z
dc.date.issued2019
dc.description.abstractA mono-energetic X-ray beam from a portable X-ray fluorescence device was used to excite characteristic X-rays from zinc in a series of nail clipping phantoms. Twenty nail clipping phantoms having equal zinc concentrations of ~40 µg/g, but with different physical characteristics, were measured individually for 300 s using a small diameter (~1 mm) X-ray beam. Energy spectra obtained from the measurements were analyzed using PyMca software. Zinc signal size varied widely between the different clippings, with a relative standard deviation of 41% observed in the combined signal from zinc Kα and Kβ characteristic X-rays. Three different normalization approaches were introduced to account for variation in the amounts of sample interrogated by the X-ray beam. All three approaches produced similar results, and successfully reduced the relative standard deviation to between 12% and 13%. A clear trend was still observed, however, between the normalized zinc signal and the thickness of clipping measured. To account for this effect, normalized signals were adjusted to calculate “thickness-corrected” values. The relative standard deviation of these thickness-corrected values was 6.2%. Reproducibility of measurement from individual clippings was excellent, with relative standard deviations on the order of 1%, with or without normalization. Overall, this new method of measuring zinc in nail shows promise for the assessment of zinc status in humans using a portable device. The method is sensitive, rapid, and requires only a single nail clipping.
dc.identifier.doi10.1016/j.apradiso.2018.12.033
dc.identifier.issn0969-8043
dc.identifier.urihttps://hdl.handle.net/20.500.14662/752
dc.identifier.urihttps://doi.org/10.1016/j.apradiso.2018.12.033
dc.language.isoen
dc.publisherElsevier
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internationalen
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.titleAssessing zinc from a nail clipping using mono-energetic portable X-ray fluorescence
dc.typeArticle

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Fleming_2019_XOS_Zn_clippings.pdf
Size:
784.52 KB
Format:
Adobe Portable Document Format

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed to upon submission
Description:

Collections